A One-Time SBOM Falls Short of the FDA's 2026 Cybersecurity Guidance
| Interlynk

The FDA's February 2026 cybersecurity guidance added no new SBOM fields. It recommends that manufacturers keep the SBOM current as the device software changes. Here is what that means for medical device makers.
Overview
The FDA reissued its medical device cybersecurity guidance on February 3, 2026. The update aligns the guidance with the new Quality Management System Regulation (QMSR) and adds no new SBOM fields. One point still matters for device makers: the FDA recommends keeping the SBOM current as the software changes.
The guidance says an SBOM, or an equivalent capability, should be kept under device configuration management. It should be updated to reflect changes to software in marketed devices. It should also support the Design and Development Files and the Medical Device File. An SBOM generated once and never updated falls behind the device. This post covers what changed, what the FDA expects, and what to do about it.
Key Facts
The FDA reissued its premarket cybersecurity guidance on February 3, 2026. It replaces the June 2025 version and aligns the guidance with the QMSR, which took effect February 2, 2026.
The update added no new SBOM fields. It updated references to the old quality system rules to reflect the QMSR and ISO 13485:2016.
Section 524B of the FD&C Act has required an SBOM in cyber device premarket submissions since March 29, 2023.
The guidance says an SBOM, or an equivalent capability, should be kept under device configuration management and updated when a shipped device's software changes.
The SBOM should support the Design and Development Files (ISO 13485 7.3.10) and the Medical Device File (4.2.3).
SBOMs should be machine-readable and follow the NTIA minimum elements. Each component should also carry its support level and end-of-support date, in the SBOM or a separate addendum. Known vulnerabilities and their risk assessments are a separate submission element.
For a 510(k), an eSTAR submission can be placed on technical screening hold for inaccurate responses or missing or irrelevant attachments, including in its Cybersecurity section.
What Changed in the February 2026 Update
The QMSR took effect February 2, 2026. It replaced the older Quality System Regulation and brought ISO 13485:2016 into 21 CFR Part 820. The FDA updated its cybersecurity guidance to reflect the QMSR and its incorporation of ISO 13485:2016. The guidance now ties cybersecurity work to design and development, software validation, risk management, complaint handling, CAPA, and postmarket activities.
The SBOM rules did not change. The update added no new SBOM fields and no new maintenance requirement. The June 2025 guidance already asked manufacturers to keep SBOMs under configuration management and update them when software changes. The 2026 version kept that and lined the language up with the QMSR.
What the FDA Expects Manufacturers to Do With the SBOM
The guidance asks for three things.
Configuration management. Keep the SBOM, or an equivalent capability, under the device's configuration management. You need to know which software components are in the device and which device version each SBOM describes.
Updates. Update the SBOM when the software in a shipped device changes. An SBOM built for the first submission stops matching the device after a patch or a dependency bump.
Quality-system documentation. The SBOM should support records like the Design and Development Files (ISO 13485 7.3.10) and the Medical Device File (4.2.3).
The FDA does not call every SBOM a standalone quality-system record. It expects the SBOM kept under configuration management and used to support that documentation. In practice, keep the SBOM current so it keeps describing the device.
Where a One-Time SBOM Breaks Down
An SBOM built for a submission can carry the information you need that day. It will not help you manage the device over time if nobody updates it. The trouble shows up in three places.
Traceability. The SBOM has to describe the software that is actually in the device. A package manifest lists the dependencies a project declares. It can miss software compiled into the firmware, linked from a library, or pulled in through a vendor SDK. Medical devices are full of these: a statically linked crypto library, a bundled real-time operating system, a vendored networking stack, an SDK or board support package from a silicon vendor. Leave them out and the SBOM no longer describes what shipped. This is the same gap that makes a supplier's SBOM hard to trust when you cannot see how it was built.
Change control. Software changes over time. Teams update libraries, fix bugs, patch security holes, and revise firmware. The SBOM has to move with it. If engineering ships an update and the SBOM stays put, the record and the device stop matching, and the quality system cannot tell.
Postmarket vulnerability management. A new vulnerability can land on a component in a device that already shipped. You need to know which devices contain it, judge the impact, and decide what to do. A current SBOM answers the first question in minutes. A file from months or years ago may not describe what is in the field anymore.
What the FDA Expects in the SBOM
The guidance starts from the NTIA minimum elements. For each component, it asks for two more things: the support level the component's maker provides through monitoring and maintenance (for example actively maintained, no longer maintained, or abandoned) and the component's end-of-support date. These can go in the SBOM or a separate addendum. SBOMs should be machine-readable, in a common format like CycloneDX or SPDX.
Known vulnerabilities sit outside the SBOM fields. The guidance asks manufacturers to identify known vulnerabilities in the device and its components, including entries in CISA's Known Exploited Vulnerabilities catalog, and to give a safety and security risk assessment and the controls for each. The SBOM points to the affected components. The vulnerability assessment says what they mean for the device. Running the SBOM through a quality check first catches the gaps that would otherwise stall both jobs.
What Section 524B Requires
Section 524B took effect March 29, 2023. It defines a cyber device as a device that includes software validated, installed, or authorized by the sponsor as a device or in a device, can connect to the internet, and contains technological characteristics, validated, installed, or authorized by the sponsor, that could be vulnerable to cybersecurity threats. For those devices, the law requires the premarket submission to include a plan to monitor, identify, and address postmarket vulnerabilities, processes to keep the device secure through updates and patches, and an SBOM covering commercial, open-source, and off-the-shelf software components.
The FDA expects a sponsor to put the Section 524B information in the submission. Its earlier refuse-to-accept policy for cyber devices expired on October 1, 2023. For a 510(k), an eSTAR submission can be placed on technical screening hold if responses are inaccurate or attachments are missing or irrelevant. That includes the Cybersecurity section. The guidance also recommends SBOM documentation for other in-scope devices that contain software, firmware, or programmable logic, including ones that do not connect to the internet. Check whether Section 524B applies to your device and follow the right guidance for the submission.
What Medical Device Makers Should Do Now
Five practical steps.
Generate the SBOM from the build. Capture the software in the finished device. For embedded systems, that means using build, compiler, linker, and firmware data. Source scans and package manifests can miss components that end up in the final firmware. This ties into what an SBOM can and cannot catch in real supply-chain risk.
Update the SBOM with every release. Tie SBOM generation to the build and release process, and keep a copy for each released version so you can tell what software is in devices already in the field.
Check SBOM quality. Look for missing components, thin metadata, and empty required fields. sbomqs can score an SBOM against established elements.
Connect the SBOM to vulnerability management. Use it to find components hit by new vulnerabilities, judge whether each one reaches the device, and record the call. VEX records and shares whether a product is affected or not affected by a vulnerability, or still under investigation, and lets you reuse earlier analysis. The FDA is clear that VEX on its own does not replace the vulnerability assessment: a sponsor still has to weigh the risk to the device and the wider system, including possible vulnerability chaining.
Keep the SBOM with your quality records. Hold it under device configuration management, tied to the Design and Development Files and the Medical Device File, so the SBOM for any device version is easy to find and its changes are easy to follow.
Where Interlynk Fits
Interlynk covers three of these parts.
lynkctl builds SBOMs from the build itself for C, C++, and embedded firmware projects. It reads build information to catch components that source scans and package manifests miss, like statically linked libraries and bundled firmware. That is what makes the SBOM match the software in the shipped device.
sbomqs scores SBOM quality on completeness and required fields across CycloneDX and SPDX. It is open source, and ENISA's October 2026 SBOM technical advisory lists it as a quality-scoring tool. Use it to check an SBOM before you rely on it.
The platform takes in the SBOM, evaluates it, assesses component risk, handles VEX, and keeps the record current across products and releases. When a vulnerability lands, you can tell which products contain the affected component.
Interlynk handles SBOM generation and management. It does not prepare or clear an FDA submission, and using the platform does not make a company FDA-compliant on its own. The quality system and the submission stay yours. We help you build and maintain the SBOMs behind them.
Bottom Line
The February 2026 update left the SBOM fields alone and carried an existing expectation into the QMSR: keep the SBOM under configuration management, update it as software changes, and use it to support the Design and Development Files and the Medical Device File. A one-time SBOM cannot do that. The work is knowing what software is in each shipped device, keeping that current, and tying it to vulnerability management.
FAQ
Did the FDA change its SBOM requirements in 2026? No new SBOM field. The February 3, 2026 update lined the guidance up with the QMSR. It recommends keeping the SBOM under device configuration management and updating it as software changes, which the June 2025 guidance already did.
Does the FDA require an SBOM for every medical device? Section 524B requires one for devices that meet the legal definition of a cyber device. The guidance also recommends SBOM documentation for in-scope devices with software, firmware, or programmable logic, including some that do not connect to the internet.
What should a medical device SBOM contain? The NTIA minimum elements, plus each component's support level and end-of-support date, in a machine-readable format like CycloneDX or SPDX. Known vulnerabilities and their risk assessments are a separate submission element.
Does the 2026 guidance make the SBOM a quality-system record? The FDA does not call every SBOM a standalone quality-system record. It says an SBOM, or an equivalent capability, should be kept under device configuration management and should support records like the Design and Development Files and the Medical Device File.
How often should an SBOM be updated? The FDA recommends updating it as the software in a shipped device changes. Doing that as part of the build and release process is the practical way to keep up.
What makes a medical device SBOM audit-ready? It describes the software in a specific device version, carries the required component details, stays current as the software changes, and connects to the quality and vulnerability records.
Sources
FDA, Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions (premarket guidance): https://www.fda.gov/media/119933/download
FDA, Quality Management System Regulation (QMSR): https://www.fda.gov/medical-devices/postmarket-requirements-devices/quality-management-system-regulation-qmsr
FDA, Cybersecurity in Medical Devices Frequently Asked Questions: https://www.fda.gov/medical-devices/digital-health-center-excellence/cybersecurity-medical-devices-frequently-asked-questions-faqs
FDA, Cybersecurity in Medical Devices: Refuse to Accept Policy for Cyber Devices: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/cybersecurity-medical-devices-refuse-accept-policy-cyber-devices-and-related-systems
FDA, Premarket Cybersecurity Guidance (June 2025 version, for the prior SBOM lifecycle language): https://www.hhs.gov/guidance/sites/default/files/hhs-guidance-documents/FDA/GUI00001825-final-PremarketCybersecurity-2025.pdf